| Model Number |
MT22608-BLU |
| Medical Device Problem Code |
Defective Alarm (1014)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/27/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
Dexcom was made aware on (b)(4) 2018 that on (b)(6) 2018, the receiver had no vibration.No additional event or patient information is available.The receiver has been received for evaluation.A follow up report will be submitted when the evaluation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
The receiver was returned for evaluation.An external visual inspection was performed and passed.The receiver was charged and will boot.The receiver data log was downloaded and reviewed, and no issues were observed related to the complaint.A global receiver functional test was performed and resulted in no failures related to the reported event.A manual "try-it" test was performed and passed.A global communication tool software test was performed and passed.A vibrator intermittent test was performed and passed.An internal visual inspection was performed and passed.Vibrator resistance was measured and found to be within specification.The reported event of no vibration was not confirmed.A probable cause could not be determined.
|
| |
|