MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON MB PATELLA PA A35; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON MB PATELLA PA A35; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5554L350
Medical Device Problem Codes Material Deformation (2976); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Pain (1994); Injury (2348)
Date of Event 07/23/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the product history records indicate devices were manufactured and accepted into final stock with no relevant reported relevant discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
It was reported that patient's left knee was revised.As reported by rep: "pain and clicking in anterior knee".An a35 patella and 5/9 cr insert were revised to an a38 x3 patella and 5/11 cr insert.
 
Event or Problem Description
It was reported that patient's left knee was revised.As reported by rep: "pain and clicking in anterior knee".An a35 patella and 5/9 cr insert were revised to an a38 x3 patella and 5/11 cr insert.
 
Additional Manufacturer Narrative
An event regarding wear, pain and audible noise involving a triathalon patella was reported.The wear was confirmed however, the pain and audible noise was not confirmed.Method & results: -device evaluation and results: material analysis, functional and dimensional inspections could not be performed as the device was not returned.Visual inspection: the reported device was not returned however photographs were provided for review.The photographs shows a recently explanted patella along with the presence of biological matter.From the photographs provided there is evidence of wear of the device.-medical records received and evaluation: rejected for medical review: "need revision operative notes, patient clinical notes, histopathology, serial x-rays, and examination of the explanted components." -device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the reported lot.Conclusion: the exact cause of the event could not be determined because insufficient information was provided.Additional information, including 'revision operative notes, patient clinical notes, histopathology, serial x-rays, and examination of the explanted components' are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON MB PATELLA PA A35
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key7797876
Report Number0002249697-2018-02577
Device Sequence Number2061587
Product Code MBH
UDI-Device Identifier07613327049909
UDI-Public07613327049909
Combination Product (Y/N)N
Initial Reporter StateDE
Initial Reporter CountryUS
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2014
Device Catalogue Number5554L350
Device Lot NumberSK4RK
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/23/2018
Supplement Date Received by Manufacturer09/12/2018
Initial Report FDA Received Date08/20/2018
Supplement Report FDA Received Date10/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age63 YR
Patient Weight90
-
-