| Medical Device Problem Code |
Pumping Problem (3016)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, the reporter contacted animas, alleging a history/settings (history/settings issue) issue; the basal history allegedly does not match the programmed basal rates.There is no indication the alleged product issue caused or contributed to an adverse event.This issue is being reported as an issue with the pump not performing as intended with regards to the history or the settings may result in over or under delivery.
|
| |
|
Additional Manufacturer Narrative
|
|
Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.Device evaluation: the device has been returned and evaluated by product analysis on 20-sep-2018 with the following findings: the black box for (b)(6)-2018 appears to be inaccurate due to a manual suspend at 01:18 followed by a manual time change to 13:34 then pump was manually resumed.Three manual suspends and manual resumes followed within three minutes, followed by a manual suspend at 13:37 and manual resume at 13:55.On investigation, the pump powered on normally and ez-prime steps were successfully completed.All electrical current readings were within specification.The pump was exercised for 24 hours without error, alarm, warning, interruption in power or overheating.There was no damage, defect or contamination of the pump¿s interior components.Investigation did not duplicate the complaint.
|
| |
|
Search Alerts/Recalls
|