Model Number 24703
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/02/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).
 
Event or Problem Description
It was reported that balloon rupture occurred.The target lesion was located in a vessel in a lower limb.A 2.5mm x 150mm x 150cm sterling sl balloon catheter was advanced for dilatation.However, during inflation, the balloon ruptured.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.
 
Additional Manufacturer Narrative
(b)(6).
 
Event or Problem Description
It was reported that balloon rupture occurred.The target lesion was located in a vessel in a lower limb.A 2.5mm x 150mm x 150cm sterling sl balloon catheter was advanced for dilatation.However, during inflation, the balloon ruptured.The procedure was completed with another of the same device.No patient complications were reported and the patient status was stable.Returned product consisted of a sterling sl balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed two buckling sections on the inner lumen in the balloon.They are located at 14.5cm and another at 15cm from the tip.The microscopic examination revealed the tip is damaged and there is a pinhole in the inner lumen located at the 15cm buckling section.Blood is visible in the inflation lumen and balloon.Inspection of the remainder of the device presented no other damage or irregularities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERLING SL
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key7798036
Report Number2134265-2018-60291
Device Sequence Number19118444
Product Code DQY
UDI-Device Identifier08714729782278
UDI-Public08714729782278
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K093720
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2020
Device Model Number24703
Device Catalogue Number24703
Device Lot Number0021685685
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/27/2018
Initial Date Received by Manufacturer 07/30/2018
Supplement Date Received by Manufacturer09/13/2018
Initial Report FDA Received Date08/20/2018
Supplement Report FDA Received Date10/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-