| Model Number |
CRPLUS |
| Medical Device Problem Code |
No Audible Prompt/Feedback (2282)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
07/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Physio-control continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device did not provide any voice prompts when the device was powered on.Without voice prompts, a lay user would not receive the audible instructions on how to properly use the device on a patient.There was no patient use associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Physio-control evaluated the customer¿s device and verified the reported issue.Physio determined that the cause of the reported issue was that the lid reed switch was remaining shorted when the lid of the device was opened.This caused the device to appear as if it was not completing a boot up cycle.Physio then replaced the lid reed switch per technical service update (tsu) 356 and after observing proper device operation through functional and performance testing the unit was returned to the customer for use.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device did not provide any voice prompts when the device was powered on.Without voice prompts, a lay user would not receive the audible instructions on how to properly use the device on a patient.There was no patient use associated with the reported event.
|
| |
|