| Catalog Number |
5537-G-519 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); Pain (1994); Swelling (2091); Injury (2348)
|
| Date of Event |
05/12/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been one other similar event for the lot and sterile lot referenced.The event relates to infection.A review of both the sterilization data and microbiology data was performed for this sterile lot commonality.It was identified that the sterilization data was within specification and no microbiology excursions were noted for this lot and as such no further review is required.It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.The following devices were also listed in this report: tri ts baseplate size 5; 5521-b-500; upgea; tri ts femur sz5 left; 5512-f-501; mwen; triathlon sym cone aug sz e; 5549-a-150; a7jh; triathlon femoral distal augment 5mm - size 5 left; 5540-a-501; koak; tri press-fit stem 18mm x 100mm; 5565-s-018; 0038666e; tri press-fit stem 15mm x 100mm; 5565-s-015; 0039945d.It cannot be determined which, if any of these devices may have caused or contributed to the patient's experience.Not returned.
|
| |
|
Event or Problem Description
|
|
The patient is a subject in stryker's triathlon tritanium cone augments outcomes study.According to the hospital: "the patient started to develop swelling and pain in the left knee on the first week of (b)(6) 2018.There was no associated causative event.The patient was admitted to the hospital on (b)(6) 2018 and underwent irrigation and debridement on (b)(6) 2018 to treat the infection that was diagnosed." the tibial insert was explanted (b)(6) 2018.
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding infection involving a triathlon insert was reported.The event was confirmed through clinician review of the medical records provided.Method & results: product evaluation and results: not performed as no product was returned for evaluation.Clinician review: a review of the provided medical records by a clinical consultant indicated: a knee arthroplasty infection developed at 2-years post previous revision surgery to require a surgical procedure with irrigation and debridement with standard bearing exchange for surgical access reasons and to improve the success rate of the procedure.Product history review: indicated all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there has been one other similar event for the lot referenced.There has been one other similar event for the sterile lot referenced.Conclusions: a review of the provided medical records by a clinical consultant indicated: a knee arthroplasty infection developed at 2-years post previous revision surgery to require a surgical procedure with irrigation and debridement with standard bearing exchange for surgical access reasons and to improve the success rate of the procedure.Arthroplasty infection is a procedure-related complication of any arthroplasty procedure with sometimes additional patient-related factors to increase infection risk.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Event or Problem Description
|
|
The patient is a subject in stryker's triathlon tritanium cone augments outcomes study.According to the hospital: "the patient started to develop swelling and pain in the left knee on the first week of may, 2018.There was no associated causative event.The patient was admitted to the hospital on (b)(6) 2018 and underwent irrigation and debridement on (b)(6) 2018 to treat the infection that was diagnosed." the tibial insert was explanted on (b)(6) 2018.
|
| |
|