| Model Number |
8888119369P |
| Medical Device Problem Code |
Migration or Expulsion of Device (1395)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/28/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during treatment, the device fell out.It was reported there was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection noted: that the catheter appeared intact.A functional evaluation found that the catheter was submerged into a water bath.The ends were clamped, and a syringe was used to inject air to observe leakage.No air bubbles were present.Both extensions were tested with acceptable results.The nodules on the silver sleeve are a normal part of the function of the device when the silver has been dissolved.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.This complaint will be used for tracking and trending purposes.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during dialysis treatment, the device fell out.It was reported the device was secured.And there were nodules on the silver ion sleeve.Customer reported the catheter was removed and another one was inserted.There was no patient injury reported.
|
| |
|
Search Alerts/Recalls
|