Catalog Number 364951
Medical Device Problem Codes Short Fill (1575); Volume Accuracy Problem (1675)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that during use of the bd vacutainer® c&s preservative urine tube, that they are not filling properly when pulling from the foley catheter.There was no report of injury or further medical intervention.
 
Event or Problem Description
It was reported that during use of the bd vacutainer® c&s preservative urine tube, that they are not filling properly when pulling from the foley catheter.There was no report of injury or further medical intervention.
 
Additional Manufacturer Narrative
Due to an it issue beginning on 7/3/2018, previously filed emdrs did not contain required fields.This supplemental emdr is filed to provide the following omitted fields.
 
Event or Problem Description
It was reported that during use of the bd vacutainer® c&s preservative urine tube, that they are not filling properly when pulling from the foley catheter.There was no report of injury or further medical intervention.
 
Additional Manufacturer Narrative
Investigation summary: bd received samples from the customer facility for investigation.The samples were evaluated and the customer's indicated failure mode for underfill with the incident lot was not observed.Evaluation and testing of the customer samples was performed and underfill was not observed.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Bd has initiated further investigation relating to this issue through a capa.The investigation is still on-going and improvements will be made as the potential causes of this issue are identified.Based on evaluation of the customer samples, the customer¿s indicated failure mode for underfill with the incident lot was not observed.Further investigation has been initiated through a capa.The investigation is still on-going and improvements will be made as the potential causes are identified.A capa has been initiated to document further investigation and root cause analysis relating to this issue.The investigation is currently on-going and will be updated as the potential root cause(s) are identified.Based on an assessment of severity and frequency, it was determined that a capa is required at this time in order to determine the root cause associated with this issue and implement corrective actions.The investigation is currently on-going and further improvements will be made as the potential root cause(s) are identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® C&S PRESERVATIVE URINE TUBE
Common Device Name
TRANSPORT CULTURE MEDIUM
Manufacturer (Section D)
BECTON, DICKINSON & CO.
150 south 1st avenue
broken bow NE 68822
MDR Report Key7798829
Report Number1917413-2018-03288
Device Sequence Number1758352
Product Code JTW
UDI-Device Identifier50382903649513
UDI-Public50382903649513
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K024240
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date05/31/2019
Device Catalogue Number364951
Device Lot Number7305995
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/10/2018
Initial Date Received by Manufacturer 08/02/2018
Supplement Date Received by Manufacturer08/02/2018
08/02/2018
Initial Report FDA Received Date08/20/2018
Supplement Report FDA Received Date09/05/2018
10/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-