| Model Number |
10601 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/25/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Age at time of event: 18 years or older. device is a combination product.
|
| |
|
Event or Problem Description
|
|
It was reported that stent dislodgment occurred.Vascular access was obtained via radial approach through a 6fr 3.5 non-bsc guide catheter and a samurai guide wire.The target lesion was located in the left anterior descending artery (lad).The physician already stented most of the lad.However, while attempting to advance a 2.50x12mm synergy drug-eluting stent through the distal lad, the non-bsc guide extension catheter stripped the stent of the delivery catheter and the device migrated to the distal lad.Subsequently, the physician advanced a 2.5x16mm synergy stent and stented the migrated stent against the wall of the distal lad.No patient complication were reported and the patient's status was stable.
|
| |
|