| Model Number |
DM0008FAA |
| Medical Device Problem Code |
Device Remains Activated (1525)
|
| Health Effect - Clinical Code |
Perforation (2001)
|
| Date of Event |
07/19/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Report inclusive.No evaluation was performed as the device was not returned.This dm0010faa perforator with lot number 463/17 was m anufactured by micromar.Device history record was reviewed by micromar and per their report, all the processes were executed according to what was planned and validated.There were no observations or nonconformities during the manufacturing process.Multiple warnings are included in the easydrill cranial perforator ifu manual including: it is essential to keep the drill perpendicular (90°) at the predetermined point of the skull to be drilled, as an excessive deviation from perpendicularity may cause the product to fail and lead to serious patient injury.Select a drill bit suitable for the bone thickness.This prevents the drill from tearing the bone or brain tissue (similar effect when the bit is not positioned at 90°).If the components of the easydrill cranial perforator become loose during trephination, discontinue use.The drill can cut or tear the dura mater when it unlocks.Check some conditions before performing the procedure, such as the existence of adhering dura mater or other anomalies adjacent to the drilling site.We will continue to monitor this complaint type for trends.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that during borehole tripanning, the drill did not stop automatically.It was reported that there was injury to the patient dura.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: patient is reported to currently have no further injury and doing fine.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Report not confirmed.Evaluation could not reproduce the reported malfunction of continues to run.No other anomalies were identified.This dm0010faa easydrill cranial perforator with lot number 463/17 was manufactured by micromar.Device history record was reviewed by micromar and per their report, all the processes were executed according to what was planned and validated.There were no observations or nonconformities during the manufacturing process was reviewed.Multiple warnings are included in the easydrill cranial perforator ifu manual including: it is essential to keep the drill perpendicular (90°) at the predetermined point of the skull to be drilled, as an excessive deviation from perpendicularity may cause the product to fail and lead to serious patient injury.Select a drill bit suitable for the bone thickness.This prevents the drill from tearing the bone or brain tissue (similar effect when the bit is not positioned at 90°).If the components of the easydrill cranial perforator become loose during trephination, discontinue use.The drill can cut or tear the dura mater when it unlocks.Check some conditions before performing the procedure, such as the existence of adhering dura mater or other anomalies adjacent to the drilling site.We will continue to track and trend this complaint type.If information is provided in the future, a supplemental report will be issued.
|
| |
|