SMITH AND NEPHEW, INC. NONRIM SPEED PIN 65 STERILE; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
| Catalog Number |
74013480 |
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
07/27/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that tip of speed pin (5mm from tip) broke and remained in femoral.It was found during x-ray check immediate post-operation.The surgeon decided that broken tip left in patient because it was located difficult region.
|
| |
|
Additional Manufacturer Narrative
|
|
The affected 6.5 mm non rimmed threaded speed pin was not returned for evaluation.However, a picture was provided.Visual inspection of the photo confirmed the tip of pin has broken off.Our investigation included a review of complaint history did not reveal additional complaints for this failure mode.A review of the manufacturing records for the listed batch which did not reveal any deviation from the standard manufacturing processes.A clinical evaluation noted that no relevant supporting clinical information has been provided to assist with a clinical investigation, and the patient's current condition is unknown.Therefore based on insufficient information, a clinical assessment cannot be performed at this time.Our investigation could not determine a specific cause of the failure.Without the return of the actual product involved, our investigation of this report is inconclusive.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should the device or additional information be received, the complaint will be reopened.
|
| |
|
Search Alerts/Recalls
|
|
|