| Model Number |
15 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
08/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Physio-control continues to investigate the reported issue and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that intermittently their device would have a blank display.When this occurred the device buttons were unresponsive.In this state defibrillation therapy may not be available to a patient if needed.There was no patient use reported for this event.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that intermittently their device would have a blank display.When this occurred the device buttons were unresponsive.In this state defibrillation therapy may not be available to a patient if needed.There was no patient use reported for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Physio-control evaluated the customer¿s device and was unable to duplicate the reported issue.The cause of the reported issued could not be conclusively determined.The customer was sent a replacement device.
|
| |
|