Catalog Number 1012530-40
Medical Device Problem Code Device Contaminated During Manufacture or Shipping (2969)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 07/09/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.Evaluation summary: visual inspection and fourier transform infrared spectroscopy (ftir) analysis was performed on the returned device.The reported complaint was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history of the reported lot revealed no other incidents.The investigation determined the loose material in the sealed header bag appears to be related to a potential product quality issue.The issue is being addressed per internal operating procedures.Abbott vascular will continue to trend the performance of these devices.
 
Event or Problem Description
It was reported that during incoming inspection of an 8x40mm absolute pro self-expanding stent system (sess) a small piece of white foreign particle was inside the sterile pouch.The incident occurred during an inspection of a consignment return at the distribution center.There was no damage to the inner pouch or the sess and the sterility was not compromised.The sess was not used.There was no patient involvement.No additional information was provided.Return device analysis revealed that the foreign material was a manufacturing component.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM
Common Device Name
SELF EXPANDING STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7799581
Report Number2024168-2018-06391
Device Sequence Number1510557
Product Code NIO
UDI-Device Identifier08717648175701
UDI-Public08717648175701
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P110028
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 08/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2019
Device Catalogue Number1012530-40
Device Lot Number6080961
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/30/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/13/2018
Initial Report FDA Received Date08/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-