| Model Number |
MMT-1755K |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/21/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
It was reported that the insulin pump alarmed critical pump error.Blood glucose was unknown at time of incident.Customer stated that the insulin pump was exposed to moisture.Troubleshooting was completed.Customer was advised to discontinue use of the insulin pump and revert to a back-up plan per healthcare provider's instruction.The customer will receive a replacement insulin pump and will return the one they currently use for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Insulin pump was received with critical pump error (open book image) alarm during testing due to moisture damage on electronic assembly.Unable to verify broken force sensor alarm due to moisture damage on electronic assembly.
|
| |
|