| Model Number |
U3575615 |
| Medical Device Problem Codes |
Retraction Problem (1536); Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/24/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No device, no medical records, or no medical images were provided to the manufacturer.The lot number for the device was provided.The device history records are currently under review.The investigation of the reported event is currently underway.Expiry date: 11/2020.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
| |
|
Event or Problem Description
|
|
It was reported that during an angioplasty procedure in the iliac artery, using an ipsilateral approach, the pta balloon allegedly ruptured at 14atm on the first inflation attempt.It was further reported that there was difficulty retracting the balloon through the sheath and subsequently the balloon and sheath were removed as one unit.Another balloon was used to complete the procedure.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Investigation summary: the device was returned for evaluation.A visual inspection found stretching in the catheter shaft, indicating retraction issues.Therefore, the retraction issues are confirmed.However, no obvious signs of rupture were noted in the balloon material.An attempt was made to inflate the balloon, however dried contrast was blocking the inflation lumen preventing the balloon from filling with water.Therefore, the investigation is inconclusive for the reported balloon rupture.It is possible that a balloon rupture could have contributed to the retraction issue through the sheath.However, the root cause for the reported balloon rupture or identified retraction issue could not be determined based upon the available information.It is unknown if procedural issues or the original sheath used in the procedure contributed to the reported event.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.(expiry date: 11/2020).
|
| |
|
Event or Problem Description
|
|
It was reported that during an angioplasty procedure in the iliac artery, using an ipsilateral approach, the pta balloon allegedly ruptured at 14atm on the first inflation attempt.It was further reported that there was difficulty retracting the balloon through the sheath and subsequently the balloon and sheath were removed as one unit.Another balloon was used to complete the procedure.There was no reported patient injury.
|
| |
|