| Model Number |
MMT-523LNAH |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/28/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump had a blank display.The customer¿s blood glucose level was 180 mg/dl at the time of the incident.Troubleshooting was performed and the display did not return.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed displacement test, self test, off no power test and all operating currents tested within the specification range.No blank display were noted.This mdr related to the (b)(4) manufacturing site has been assigned a medwatch number from the medtronic minimed (b)(4) site, per variance 5.
|
| |
|