| Catalog Number |
03183734190 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/31/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This event occurred in (b)(6).
|
| |
|
Event or Problem Description
|
|
The customer received questionable tp2 total protein gen.2 results for one patient sample from cobas 6000 c 501 module serial number (b)(4).The initial result was 9.0 g/dl and was reported outside of the laboratory.The repeat result with a 1:10 dilution was 9.8 g/dl.The result by shikoku chuken method was 7.7 g/dl.There was no allegation of an adverse event.As calibration and qc data were acceptable, a general reagent or application could be ruled out.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation did not identify a product problem. the cause of the event could not be determined. .
|
| |
|