Model Number MI1200 SYNCHRONY PIN
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).(exemption number e2015033).The device has not been explanted.If it should be explanted, it is to be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Event or Problem Description
The recipient's hearing performance with the device has decreased between 6 and 12 months post-activation.He reports intermittencies in sound and his ability to understand speech fluctuates.No incident of accident or trauma is reported.Re-implantation has been scheduled.
 
Additional Manufacturer Narrative
Med-el elektromedizinische geraete gmbh and vibrant med-el submit mdr reports on behalf of med-el corporation (exemption number e2015033).Conclusion: damage to the active electrode, likely caused by minute device mobility, was determined to have led to device failure over time.The investigation results appear to match the problems mentioned in the recipient report.This is final report.
 
Event or Problem Description
The recipient's hearing performance with the device has decreased between 6 and 12 months post-activation.He reports intermittencies in sound and his ability to understand speech fluctuates.No incident of accident or trauma is reported.The recipient was re-implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM
Common Device Name
COCHLEAR IMPLANT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
MDR Report Key7800539
Report Number9710014-2018-00648
Device Sequence Number228840
Product Code MCM
UDI-Device Identifier09008737094010
UDI-Public(01)09008737094010
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P000025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMI1200 SYNCHRONY PIN
Device Catalogue Number31090
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/26/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 08/14/2018
Supplement Date Received by Manufacturer08/14/2018
Initial Report FDA Received Date08/21/2018
Supplement Report FDA Received Date11/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
-
-