DEPUY ORTHOPAEDICS INC US ALTRX NEUT 32IDX56OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS Back to Search Results
Catalog Number 122132056
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Not Applicable (3189)
Date of Event 07/29/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the poly liner became dislodged from cup.Patient stated that he was dancing and felt a ¿pop¿ in his hip and knew something was not right.Patient went to the emergency room, after x-rays and ct scan it was verified that the liner had dislodged from the cup.The surgeon did not know if the liner was defective or if there was impingement that could have caused the separation of the liner.Surgeon revised the acetabular cup because he felt the position of the cup was not optimal.Doi: (b)(6) 2017; dor: (b)(6) 2018; right hip.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide complaint database search found no other reported incident(s) against the provided product/lot combination(s) since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTRX NEUT 32IDX56OD
Common Device Name
PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key7800647
Report Number1818910-2018-67202
Device Sequence Number14044674
Product Code LPH
UDI-Device Identifier10603295016137
UDI-Public10603295016137
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K072963
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number122132056
Device Lot NumberH09202
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/29/2018
Supplement Date Received by Manufacturer09/19/2018
Initial Report FDA Received Date08/21/2018
Supplement Report FDA Received Date09/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age59 YR
-
-