| Model Number |
9-ASD-024 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/20/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The reported event of deformity upon deployment could not be confirmed.The investigation confirmed the device met all functional specifications when analyzed at abbott under non-physiological conditions.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported event remains unknown.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, a 24mm amplatzer septal occluder was selected for implant.When the left disc was deployed in the left atrium, the disc deformed.The device was re-sheathed and deployed several times with the same result.The device was removed from the patient without complication.A second 24mm device was deployed in the left atrium but the left disc came out of the sheath deformed.The second 24mm device was removed and a 26mm amplatzer septal occluder was successfully implanted.
|
| |
|