• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - MAPLE GROVE QUANTUM¿ MAVERICK¿; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - MAPLE GROVE QUANTUM¿ MAVERICK¿; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number H7493808015250
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/28/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Device evaluated by mfr: returned product consisted of a hypotube in two pieces, the shaft, balloon and tip of a quantum maverick balloon catheter; the hub was not returned for analysis.The outer shaft, inner shaft, balloon and tip were microscopically examined.The distal hypotube, shaft, balloon and tip were completely separated 50.6cm from the distal tip.81.5cm of the proximal hypotube was returned, the proximal end of the hypotube is separated and the hub was not returned.The fracture faces were oval as if kinked prior to separation.There are numerous hypotube and shaft kinks.Inspection of the remainder of the device presented no other damage or irregularities.There was no evidence of any material or manufacturing deficiencies contributing to the damage and the reported difficulty.The investigation conclusion is operational context as the product meets the design and manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
 
Event or Problem Description
Reportable based on device analysis completed on 03-aug-2018.It was reported that crossing difficulties were encountered.The target lesion was located in a coronary artery.A 2.5mm x 15mm quantum¿ maverick¿ balloon catheter was advanced but failed to cross the leison.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.However, the returned device analysis revealed hypotube break and hub detached/separated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUANTUM¿ MAVERICK¿
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
MDR Report Key7801320
Report Number2134265-2018-07363
Device Sequence Number14080388
Product Code LOX
UDI-Device Identifier08714729392767
UDI-Public08714729392767
Combination Product (Y/N)N
Initial Reporter CountryCN
PMA/510(K) Number
P860019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2019
Device Model NumberH7493808015250
Device Catalogue Number38080-1525
Device Lot Number0019783657
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/09/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/03/2018
Initial Report FDA Received Date08/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-