Catalog Number KX203
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Codes Hematoma (1884); Blood Loss (2597)
Date of Event 07/26/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient underwent balloon kyphoplasty due to vertebral compression fracture at l1.Intra-op, the balloon ruptured on the right side of l1.The balloon was deflated successfully.Blood began pulsating out of the right cannula.Then, balloon on left was deflated and inserted into the right cannula and inflated.Now, blood began pulsating out of left cannula as well.The surgeon decided to inject surgiflow down both cannulas.Bleeding stopped at this time.Both the balloons were inflated once more.Finally, about 7.5 cc of cement was injected into l1 successfully.A 300 ml of blood was lost due to this event.Ct conducted after the surgery revealed a small psoas hematoma.The patient is stable and is recovering from hematoma.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KYPHON XPANDER II INFLATABLE BONE TAMP
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
MEDTRONIC MEXICO (TIJUANA)
avenida paseo cucapah # 10510
parque industrial ei lago
tijuana,bc 22570
MX  22570
MDR Report Key7801504
Report Number9612164-2018-02110
Device Sequence Number2488952
Product Code HXG
UDI-Device Identifier00643169093676
UDI-Public00643169093676
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K101864
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberKX203
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/26/2018
Initial Report FDA Received Date08/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age79 YR
-
-