| Model Number |
BC0410C |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/24/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The lot number was provided.A review of the approved device history record indicated the lot met all release criteria.A lot history trending review was performed and there were no similar complaints for this lot number.The device was not returned to the manufacturer; therefore, a product evaluation could not be performed.The root cause is unknown.
|
| |
|
Event or Problem Description
|
|
It was reported that the balloon ruptured during the procedure.The balloon was removed from the patient without incident.There was no reported intervention or patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned for evaluation.During investigation, the balloon was noted to have a linear tear.Further inspection determined that the polyimide ring was missing.Based on the investigation and provided information, the reported complaint of a ruptured balloon was confirmed.The root cause cannot be definitively determined; however, the damage to the device is consistent with over-inflation.
|
| |
|
Search Alerts/Recalls
|