Model Number 693575
Medical Device Problem Codes Signal Artifact/Noise (1036); Failure to Capture (1081); Fracture (1260); High impedance (1291); Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574); High Capture Threshold (3266)
Health Effect - Clinical Codes Dyspnea (1816); Shock from Patient Lead(s) (3162)
Date of Event 07/13/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitant medical products: 407652 lead, implanted: (b)(6) 2005.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient was inappropriately shocked by the right ventricular (rv) lead while asymptomatic and went to the emergency room (er).The rv lead triggered a lead integrity alert (lia) due to oversensing of noise with high rate non-sustained (hr-ns) and sensing integrity counter (sic) episodes.The lead had high impedance, high thresholds with loss of capture, and a confirmed fracture.A magnet was placed over the device to prevent any further shocks and the lead was programmed off.The lead remains in the patient out of service.It was further reported that the right atrial (ra) lead has high thresholds.The lead remains in use.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the lead was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated oversensing associated with the right ventricular lead.Analysis of the device memory indicated the criteria for the right ventricular lead integrity alert were met.Analysis of the device memory indicated the impedance of the right ventricular pacing lead was beyond the expected upper range.Analysis of the device memory indicated oversensing due to non-physiologic signals/sensing integrity counter.Analysis of the device memory indicated unexpected delivery of ventricular tachyarrhythmia therapy.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was further reported that the patient later presented to the er again with shortness of breath.Upon review of the device interrogation, therapies and the rv lead were programmed off into an atrial only mode.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPRINT QUATTRO SECURE S MRI SURESCAN
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key7801940
Report Number2649622-2018-13673
Device Sequence Number228855
Product Code LWS
UDI-Device Identifier00613994446367
UDI-Public00613994446367
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P920015
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/25/2012
Device Model Number693575
Device Catalogue Number693575
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/25/2018
Supplement Date Received by Manufacturer08/21/2018
10/05/2019
Initial Report FDA Received Date08/21/2018
Supplement Report FDA Received Date09/13/2018
10/26/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0474-2011
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DDMB1D1 ICD
Outcome Attributed to Adverse Event Hospitalization; Life Threatening; Required Intervention;
Patient Age65 YR
-
-