BECTON DICKINSON BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Catalog Number 367297
Medical Device Problem Code Volume Accuracy Problem (1675)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The manufacturing location for this product is (b)(4).This site is an oem manufacturing site.Therefore, bd corporate headquarters in (b)(4) has been listed, and the (b)(4) fda registration number has been used for the manufacture report number.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that during use of the bd vacutainer® safety-lok¿ blood collection set, blood only flowed half way down the tubing."needle will give a blood flash, & progress halfway down tubing, but then stops & will not come any further.When needle is removed from vein, bubbles appear in tubing & blood leaks out of needle tip." there was no report of exposure, serious injury or medical intervention.
 
Event or Problem Description
It was reported that during use of the bd vacutainer® safety-lok¿ blood collection set, blood only flowed half way down the tubing."needle will give a blood flash, & progress halfway down tubing, but then stops & will not come any further.When needle is removed from vein, bubbles appear in tubing & blood leaks out of needle tip." there was no report of exposure, serious injury or medical intervention.
 
Additional Manufacturer Narrative
Correction: due to an it issue beginning on 7/3/2018, previously filed emdrs did not contain required fields.This supplemental emdr is filed to provide the following omitted fields: event attributed to: other, device single use?: no, device returned to manufacture: yes.Investigation summary: bd received samples from the customer facility for investigation.The samples, along with retention samples selected from bd inventory, were evaluated and the customer's indicated failure mode for clogged tubing and leakage at the needle tip with the incident lot was not observed.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Based on evaluation of the customer and retain samples, the customer¿s indicated failure mode for clogged tubing and leakage at the needle tip with the incident lot was not observed as all samples met the required specifications.Based on the investigation, a root cause could not be determined.The product was found to be in conformance and meet release specifications.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.The bd business team regularly reviews the collected data for identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® SAFETY-LOK¿ BLOOD COLLECTION SET
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
MDR Report Key7802188
Report Number2243072-2018-01116
Device Sequence Number13836970
Product Code JKA
UDI-Device Identifier50382903672979
UDI-Public50382903672979
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K991088
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date04/30/2021
Device Catalogue Number367297
Device Lot Number8D2981
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/08/2018
Initial Date Received by Manufacturer 08/01/2018
Supplement Date Received by Manufacturer08/01/2018
Initial Report FDA Received Date08/21/2018
Supplement Report FDA Received Date09/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-