Catalog Number 280135
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Event or Problem Description
A customer in (b)(6) contacted biomérieux to report intermittent lls errors in association with the nuclisens® easymag® disposable vessels used on one of their easymag® systems.The customer reported that results were delayed more than 24 hours with no impact to patient care.An internal biomérieux investigation will be initiated.
 
Additional Manufacturer Narrative
An internal investigation was performed for a customer in martinique who contacted biomérieux to report intermittent lls errors in association with the nuclisens® easymag® disposable vessels used on their easymag® systems.The customer reported obtaining several random lls errors on his two easymags with several vessel batches (z108kb / z109am / z109do and z109fn).The customer also reported solving the lls errors by switching on another disposable batch z109km.The customer sent the data for the impacted samples ( plasma, respiratory samples (throat swab, sputum), urine) which were pre-treated when necessary.The fse (field service engineer) visited the site and found no equipment issue via the lls test performed and no particular issue highlighted.The run report was analyzed by r&d and service engineering : lls errors are due to tips clogged by samples (not homogenous) and were displayed on different positions on the vessel, since lls occurred at the first aspiration, it seems to be sample related or due to crystals in the lysis buffer (occurring when the buffer is stored below the ambient temperature) the impacted samples failed as no eluate obtained.The storage conditions of the lysis buffer 1l are not clear and could be the root cause.Disposables supplier medisize was informed on (b)(6) 2018 : complaint (b)(4) opened on (b)(6) 2018.Batch documentation showed no direct link between the impacted batches.Z108kb and z109am were manufactured with old plastic raw material (moplen) whereas z109do and z109fn are made of new raw material (sabic) review of batch documentation for both assembled product and injection moulding products showed no non conformity / planned deviation / complaint related to these batches.The exact root cause is unknown, however, the most probable cause is that tips were clogged by lysis buffer crystals causing lls errors on the easymag instruments.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUCLISENS® EASYMAG® DISPOSABLE
Common Device Name
NUCLISENS® EASYMAG® DISPOSABLE
Manufacturer (Section D)
BIOMERIEUX SA
5, rue des berges
grenoble cedex, 38024
FR  38024
MDR Report Key7802499
Report Number3002769706-2018-00155
Device Sequence Number2488276
Product Code JJH
UDI-Device Identifier03573026139360
UDI-Public03573026139360
Combination Product (Y/N)N
Initial Reporter CountryMB
PMA/510(K) Number
C1 EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/28/2019
Device Catalogue Number280135
Device Lot NumberZ109FN
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/30/2018
Supplement Date Received by Manufacturer12/20/2018
Initial Report FDA Received Date08/21/2018
Supplement Report FDA Received Date01/18/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-