| Catalog Number |
06D19-68 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/04/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
All available patient information was included.Additional patient details are not available.An evaluation is in process.A final report will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The customer observed a donor generated false repeat reactive, confirmed (b)(6) prism (b)(6) results.(b)(6) was prism hbsag repeat reactive and sent to a reference lab which generated confirmed (b)(6) results id (b)(6) (plasma bag aliquot from same donor) was prism (b)(6) repeat reactive and sent to the reference lab generating nonconfirming prism (b)(6) results.Additional donor testing was nat and anti-hbc negative.No impact to donor management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The evaluation of the customer issue included a search for similar complaints, a review of historical data and a review of labelling.Upon reviewing the likely cause lots, no adverse trend, non-statistical trends, non-conformances, potential non-conformances, or deviations were identified for the customer's issue.Labeling was reviewed and found to be adequate.Testing the sample with reagent lots and reference lots interpreted repeat reactive non-confirming result.Testing was performed using in-house retained kits and the testing met all specifications.Initial reactive rate, the repeat reactive rate and the specificity of complaint lot numbers were within package insert specifications.Based on the available information no product deficiency was identified.
|
| |
|
Search Alerts/Recalls
|