| Model Number |
LF5544 |
| Medical Device Problem Codes |
Flare or Flash (2942); Noise, Audible (3273)
|
| Health Effect - Clinical Codes |
Burn(s) (1757); Partial thickness (Second Degree) Burn (2694)
|
| Date of Event |
08/03/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Occupation: health professional (hpro).If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, the device was not working and they heard a bang.A second instrument was used, but was not okay as well, because it was faulty as the tissue was burned lateral.According to the nurse with one of the instruments, while it was being used the insulation slipped off, was removed from the device, in a backward direction of the jaw and the surrounding tissue was burned.The disengaged part was removed from the device's tip and it did not fall into the patient's cavity.Localized burn of level 1-2 was caused by the non-insulated part on the jaw.Patient is fine and there were no other complications.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: two devices were received for evaluation.The returned product did not meet specification as received.Visual inspection found the clear boot was torn, all pieces were still attached and present on one of the devices.The other device had no visual defects.The customer reported that the device was not working and they heard a bang.The reported condition was not confirmed.The investigation found the devices to function normally and within specifications.The devices were recognized when plugged into the generator.The devices were activated on a saline soaked towel with acceptable results and a visual seal effect was observed.The devices functioned normally when activated in the vlmode.No abnormal noise was heard when the devices were activated.The investigation found the device to function normally and within specifications.The reported condition was confirmed.The device failed hipot testing.The investigation identified the root cause of the reported event to be user error.The instructions for use (ifu) states to prevent damage to the flexible insulation proximal to the jaws, confirm the handle is fully closed prior to insertion into and extraction from the cannula.Use the appropriately sized cannula to allow for easy insertion and extraction of the instrument.Failure to do so may impact the integrity of the flexible insulation.Cannulas with hard, non-beveled openings may cause the flexible insulation to retract, which may compromise the insulation.If retraction occurs, the instrument must be discarded.Do not attempt to clean the flexible insulation.Cleaning may damage insulation.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|