| Catalog Number |
3910-200-035 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/30/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the anchor broke.All pieces were retrieved from the patient.
|
| |
|
Event or Problem Description
|
|
It was reported that the anchor broke.All pieces were retrieved from the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Alleged failure: mrs.(b)(6) from the clinic reported to (b)(6) the following event: "an anchor ref (b)(4) that broke on impact." the failure identified in the investigation is consistent with the complaint record.The probable root causes could be: excessive force applied on device, bone too hard for anchor insertion, or drill not inserted to proper depth.
|
| |
|