| Model Number |
4400804S |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/24/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).The product was not returned for analysis.Device history record (dhr) review was conducted and the product met quality requirements for product acceptance.Additional information is pending and will be submitted within 30 days upon receipt.
|
| |
|
Event or Problem Description
|
|
During use, it was reported that a powerflex pro balloon catheter was delivered to the lesion and ruptured at four atmosphere when the physician inflated the balloon.There was no patient injury.The lesion was the external iliac artery.The device will not be returned for analysis because of infectious disease.
|
| |
|
Additional Manufacturer Narrative
|
|
During use, it was reported that an 8x40mm 80cm powerflex pro balloon catheter (bc) was delivered to the lesion and ruptured at four atmospheres (4 atm) when the physician inflated the balloon.There was no patient injury.The lesion was the external iliac artery.The device was not returned for analysis due to the patient¿s infectious disease.A device history record (dhr) review of lot 17737037 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿balloon burst at/below rbp¿ could not be confirmed as the device was not returned for analysis.The exact cause of the burst could not be determined.Based on the limited information available for review, vessel characteristics (although not provided) and procedural/handling factors may have contributed to the reported event since calcified/resistant lesions can cause damage to a balloon.According to the instructions for use, which is not intended as a mitigation, ¿the rated burst pressure is based on the result of in vitro testing.At least 99.9% of the balloons (with a 95% confidence level) will not burst at or below their rated burst pressure.Use of a pressure monitoring device is recommended to prevent over pressurization.Prior to angioplasty, the catheter should be examined to verify functionality and ensure that its size and shape are suitable for the specific procedure for which it is to be used.¿ neither the dhr review nor the information available suggests a design or manufacturing related cause for the reported event.Therefore, no corrective/preventive action will be taken at this time.
|
| |
|
Search Alerts/Recalls
|