• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. VALEO BILIARY STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. VALEO BILIARY STENT Back to Search Results
Model Number EV08561CD
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/18/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during an angioplasty procedure in the iliac artery, the stent became dislodged from the balloon prior to entering the sheath.Another device was used to complete the procedure.There was no reported patient injury.
 
Event or Problem Description
It was reported that during an angioplasty procedure in the iliac artery, the stent became dislodged from the balloon prior to entering the sheath.Another device was used to complete the procedure.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records have been reviewed with special attention to the raw materials, subassemblies, manufacturing process, and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Investigation summary: the device was returned for evaluation.A visual inspection found the stent was dislodged proximally on the balloon.Several struts on the stent were noted to be bunched.The balloon was examined under magnification and crimp marks were visible on the balloon.Therefore, based on the returned sample, the investigation is confirmed for dislodgement of the stent.The definitive root cause for the identified dislodged stent could not be determined based upon available information.It is unknown whether procedural issues involving the user's sheath contributed to the event.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALEO BILIARY STENT
Common Device Name
BILIARY STENT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7803616
Report Number2020394-2018-01516
Device Sequence Number16466353
Product Code FGE
UDI-Device Identifier00801741000850
UDI-Public(01)00801741000850
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K052132
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/17/2019
Device Model NumberEV08561CD
Device Catalogue NumberEV08561CD
Device Lot NumberGFBU2137
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/09/2018
Initial Date Received by Manufacturer 07/25/2018
Supplement Date Received by Manufacturer10/25/2018
Initial Report FDA Received Date08/21/2018
Supplement Report FDA Received Date11/01/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age72 YR
Patient Weight78
-
-