Catalog Number 5531G209
Medical Device Problem Codes Material Erosion (1214); Material Deformation (2976)
Health Effect - Clinical Code Injury (2348)
Date of Event 07/25/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history records indicates that the reported devices were manufactured and accepted into final stock with no reported discrepancies.The complaint history review indicated that there were no similar events for the reported lot.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Not returned.
 
Event or Problem Description
Dr (b)(6) from the hospital reported to mr (b)(6), rsm joint replacement the following event: "i had a call from dr (b)(6) on wednesday afternoon for an unusually worn pe x3 triathlon implant.I recovered the implant yesterday and actually there is a big wear for a 1 year bp.
 
Event or Problem Description
Dr (b)(6) from the hospital reported to mr (b)(6), rsm joint replacement the following event: "i had a call from dr (b)(6) on wednesday afternoon for an unusually worn pe x3 triathlon implant.I recovered the implant yesterday and actually there is a big wear for a 1 year bp.
 
Additional Manufacturer Narrative
An event regarding wear involving a triathlon insert was reported.The event was confirmed by material analysis of the returned device.Method & results: device evaluation and results: one insert was returned for evaluation.Burnishing, scratching and third-body indentations were observed on the insert condyles these are common damage modes of uhmwpe.A material analysis has been performed.The report concluded: damage was observed on the condyles of the femoral component, consistent with contact against a hard object.Burnishing, scratching and third-body indentations were observed on the insert condyles.These are common damage modes of uhmwpe.Eds analysis showed the femoral component and baseplate were consistent with astm f75 alloy.Biological fixation was observed on the femoral component and baseplate.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined.Medical records received and evaluation: no medical records were received for review with a clinical consultant.Device history review: device history review indicated the devices accepted into final stock from the reported lot were free from discrepancies.Complaint history review: there have been no other similar events for the reported lot.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Additional information including operative reports, progress notes, x-rays as well as patient history and follow up notes are required to complete the investigation for determining a root cause.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X3 TRIATHLON CS INSERT #2 9MM
Common Device Name
KNEE JOINT PATELLOFEMOROTIBIAL METAL/POLYMER POROUS-COATED UNCEMENTED PROSTHESIS
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key7803733
Report Number0002249697-2018-02634
Device Sequence Number16527809
Product Code MBH
UDI-Device Identifier07613327045659
UDI-Public07613327045659
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/31/2022
Device Catalogue Number5531G209
Device Lot NumberLFN559
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/20/2018
Initial Date Received by Manufacturer 07/25/2018
Supplement Date Received by Manufacturer02/15/2019
Initial Report FDA Received Date08/21/2018
Supplement Report FDA Received Date03/15/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-