| Model Number |
RT226 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
07/13/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Method: the complaint rt226 infant low flow breathing circuit was received at fisher & paykel healthcare (b)(4) for investigation.The subject circuit was visually inspected.Results: the watertrap end connector of the short expiratory tube was found disconnected.No damage was found to the connector or the tube cuff.Conclusion: we are unable to determine the cause of this complaint.All rt226 infant dual-heated evaqua2 breathing circuits are visually inspected and pressure and flow tested during production, and those that fail are rejected.Our user instructions that accompany the rt226 state the following: check all connections are tight before use.Perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient.Set appropriate ventilator alarms.
|
| |
|
Event or Problem Description
|
|
A hospital in (b)(6) reported via our distributor that an rt226 infant low flow breathing circuit failed the initial leak test of the servo-i ventilator.They also noticed an air leak from the connection between the expiratory limb and the water trap.There was no patient involvement.
|
| |
|