Catalog Number TRT10
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch # p56484.Date of event: month and day unknown; only year (2018) is known.Device evaluation: the analysis results found that the trt10 cartridge reload was received with the reload forks damaged, fully loaded with staples, and the swing tab was noted to be broken, making the reload nonfunctional.No functional test could be performed due the condition of the reload.While no conclusion could be reached as to how the damage to the device occurred, it is possible that the damaged was due to an improper handling of the device.Please refer to the "instruction for use" for more information.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
 
Event or Problem Description
It was reported that during an unknown procedure, when the surgeon tried to fire the device, the knife did not go forward.There were no patient consequences reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key7804315
Report Number3005075853-2018-12270
Device Sequence Number19199719
Product Code GDW
UDI-Device Identifier20705036003233
UDI-Public20705036003233
Combination Product (Y/N)N
PMA/510(K) Number
K020779
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2022
Device Catalogue NumberTRT10
Device Lot NumberP4RD4H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/06/2018
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/06/2018
Initial Report FDA Received Date08/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-