| Model Number |
32G X 6MM |
| Medical Device Problem Codes |
Improper or Incorrect Procedure or Method (2017); Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Information (3190)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The tests on archival sample were performed for compliance with the parameters established in product specifications.No defects or other nonconformities were found.The tests on customer sample were performed for compliance with the parameters established in product specifications.The defects under complaint were not confirmed in the sample from customer.
|
| |
|
Event or Problem Description
|
|
User is not able to use or prime insulin pen needle.Probably needles were bent.
|
| |
|
Additional Manufacturer Narrative
|
|
Method code changed.
|
| |
|
Event or Problem Description
|
|
User is not able to use or prime insulin pen needle.Probably needles were bent.
|
| |
|
Additional Manufacturer Narrative
|
|
The tests on customer sample were performed for compliance with the parameters established in product specifications.The defects under complaint were not confirmed in the sample from customer.The probable cause of such defect was non-axial twisting pen needle on pen injector.
|
| |
|
Event or Problem Description
|
|
User is not able to use or prime insulin pen needle.Probably needles were bent.
|
| |
|