• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS ANGIODYNAMICS; STOPCOCK / ARTERIAL EXTENSION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS ANGIODYNAMICS; STOPCOCK / ARTERIAL EXTENSION SET Back to Search Results
Catalog Number H965907030621
Medical Device Problem Code Component Misassembled (4004)
Health Effect - Clinical Code Not Applicable (3189)
Date of Event 07/20/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A photo has been received from the distributor which confirms the reported event.The investigation is on-going and upon completion , a supplemental medwatch will be submitted.((b)(4)).
 
Event or Problem Description
As reported by angiodynamics' distributor in (b)(4), blue non-vented port protector cap connected to stopcock end port fitting should have been a vented white port protector cap.The non-vented cap possibly compromises a sterile pathway for this device during the sterilization process.This one unit was found during 100% incoming inspection at the distributor and had not been distributed to a hospital.
 
Additional Manufacturer Narrative
A review of the device history records was performed for the reported packaging lot 5319428 for any deviations related to the reported defect of the complaint.The review confirms that the lot met all material, assembly and performance specifications.A review of the recent angiodynamics complaint report was performed for the stopcock/manifold product family and the failure mode " cap incorrectly constructed/assembled." no adverse trend was indicated.The sample was not returned to angiodynamics, however a photo was provided which confirmed the reported issue: the incorrect port protector was attached to the end port of the stopcock.The shipment of devices from lot 5319428 shipped to the reported distributor consisted on (b)(4) units.(b)(4).The root cause of the incorrect cap attachment has been determined to be a manufacturing assembly error in which responsible employees failed to properly attach the components and verify the proper attachments per the drawing and packaging op.Codes.Packages are 100% visualized for this type of non-conformance and the employees who packaged the affected lot did not identify the non-conforming unit during their inspection.In an effort to heighten awareness of the performance of our products in the field, recent complaints are reviewed in a weekly meeting with the entire department.All responsible packaging employees have been retrained on the applicable procedures.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIODYNAMICS
Common Device Name
STOPCOCK / ARTERIAL EXTENSION SET
Manufacturer (Section D)
ANGIODYNAMICS
10 glens falls technical park
glens falls NY 12801
MDR Report Key7804936
Report Number1317056-2018-00162
Device Sequence Number13820298
Product Code DQO
UDI-Device IdentifierH965907030621
UDI-PublicH965907030621
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K790082
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date04/30/2021
Device Catalogue NumberH965907030621
Device Lot Number5319428
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/20/2018
Supplement Date Received by Manufacturer07/20/2018
Initial Report FDA Received Date08/22/2018
Supplement Report FDA Received Date09/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-