Model Number 3948L
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 07/27/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a distal gastrectomy, they set the anvil to the anvil fork, and the surgeon clamped the duodenum and tried to fire the device, however, the knife did not advance.Replaced by a new reload, and the firing was completed.Reoperation was required due to the issue.The surgical time was extended by less than 30 min.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The loading unit was received partially fired with no visual abnormalities.The single-use loading unit (sulu) was loaded into a pmv representative instrument.The sulu and representative device were applied to the appropriate test media with acceptable results.A review of the device history record indicates this product was released meeting all quality release specifications at the time of manufacture.Replication of the reported condition may occur if the firing stroke is not complete and the firing knob is not fully retracted after firing or if the instrument is applied against an excessive amount of tissue during the clinical application.The root cause of the observed damage was misuse of the product which caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ILA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
MDR Report Key7805213
Report Number2647580-2018-04161
Device Sequence Number16407224
Product Code GDW
UDI-Device Identifier10884521063150
UDI-Public10884521063150
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K885047
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2022
Device Model Number3948L
Device Catalogue Number3948L
Device Lot NumberP7K1727X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/27/2018
Initial Date Received by Manufacturer 08/03/2018
Supplement Date Received by Manufacturer09/13/2018
Initial Report FDA Received Date08/22/2018
Supplement Report FDA Received Date10/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-