| Catalog Number |
750170 |
| Medical Device Problem Codes |
Fitting Problem (2183); Defective Component (2292); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/24/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).
|
| |
|
Event or Problem Description
|
|
Information was received indicating that the speaking valve was unable to connect to a smiths medical portex® bivona® aire-cuf® tracheostomy tube.It was noted that the 15mm connector is out of specification.There were no reported adverse effects.
|
| |
|
Additional Manufacturer Narrative
|
|
One 7.0mm id portex bivona aire tracheostomy tube was returned for analysis.Upon visual examination an oval deformation was noted at the connector openings; confirming complaint.The connector on the returned device was inspected and determined not to match the manufacturing specification for od and taper.A 100% iso taper inspection was implemented immediately prior to packaging the product in january 2016.No internal failures were identified since implementing the 100% iso taper inspection.Based on the evidence, the root cause is unknown.
|
| |
|