SI-BONE, INC. IFUSE IMPLANT SYSTEM; ORTHOPEDIC ROD Back to Search Results
Model Number 7070M-90
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Arthralgia (2355)
Date of Event 07/26/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Based on the information provided, review of the surgical technique manual, ifu, certificates of analysis and fmeas, there is no indication of device failure and no indication that the device was out of specification.The root cause is the surgeon's decision to use a shorter implant as a precaution.
 
Event or Problem Description
In (b)(6) 2018, the surgeon performed a left side si joint arthrodesis on the patient placing two implants along with an iliosacral screw on a pelvis trauma case.The patient complained of continued radiculopathy that he had before the procedure.Imaging confirmed that neither of the implants had breached the neuroforamen, but the surgeon decided to remove the longer implant and replace it with a shorter implant as a precaution due to the patient's history of sciatica.Three days after the initial procedure, the surgeon performed a revision procedure where he removed the 70 mm implant and replaced it with a shorter implant, 55 mm, using the explant void.No other preexisting implants were adjusted or removed.The status of the patient following the revision procedure is not known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IFUSE IMPLANT SYSTEM
Common Device Name
ORTHOPEDIC ROD
Manufacturer (Section D)
SI-BONE, INC.
471 el camino real
suite 101
santa clara CA 95050 4482
MDR Report Key7805379
Report Number3007700286-2018-00064
Device Sequence Number1536331
Product Code OUR
UDI-Device Identifier00851085007541
UDI-Public00851085007541
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K080398
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/05/2023
Device Model Number7070M-90
Device Lot Number2631391
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/26/2018
Initial Report FDA Received Date08/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
-
-