• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SCR BIORCI 8X25MM STRL BX 1; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SCR BIORCI 8X25MM STRL BX 1; SCREW, FIXATION, BONE Back to Search Results
Model Number 7207560
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/13/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).
 
Event or Problem Description
It was reported that the distal part of the screw was broke off inside a patient.All the parts were removed completely from the patient.No delay was reported and a backup was available.No patient injuries were reported.
 
Additional Manufacturer Narrative
One 8x25 mm biorci screw was returned for evaluation.Visual assessment of the screw confirmed the reported breakage.Approximately 2.5 mm of the distal threads have broken off and were not returned for examination.Dimensional assessment of the screws major diameter and wall thickness was performed and found to meet print specifications.This investigation could not identify any evidence of product contribution to the reported complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCR BIORCI 8X25MM STRL BX 1
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
MDR Report Key7806171
Report Number1219602-2018-01108
Device Sequence Number1116787
Product Code HWC
UDI-Device Identifier03596010426222
UDI-Public(01)03596010426222(17)211019(10)50626556
Combination Product (Y/N)N
Initial Reporter CountryMX
PMA/510(K) Number
K992396
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/19/2021
Device Model Number7207560
Device Catalogue Number7207560
Device Lot Number50626556
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/17/2018
Initial Date Received by Manufacturer 07/27/2018
Supplement Date Received by Manufacturer09/25/2018
Initial Report FDA Received Date08/22/2018
Supplement Report FDA Received Date09/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-