MAUDE Adverse Event Report: PHYSIO-CONTROL, INC. - 3015876 LIFEPAK CR(R) PLUS DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC. - 3015876 LIFEPAK CR(R) PLUS DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number CRPLUS
Medical Device Problem Code Power Problem (3010)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The customer was provided with a replacement device.Physio-control continues to investigate the reported issue and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
 
Event or Problem Description
The customer contacted physio-control to report that their device had all three icons (charge pak, attention and service wrench) present on its readiness display.The customer advised that the device did not provide any voice prompts when the device was powered on.The three icons being illuminated indicates that the device may not be able to provide sufficient defibrillation therapy, if necessary.There was no patient use associated with the reported event.
 
Additional Manufacturer Narrative
Physio-control evaluated the customer's device and verified the reported issue.Physio-control determined that the cause of the reported issue was that hybrid layer capacitor (hlc) batteries, designators bt1, bt2 and bt3, of the analog pcb would no longer hold a charge.The hlc batteries would no longer hold a charge because they were damaged after having previously depleted to zero (0).Excessive on time was observed and is likely due to an object being placed on top of the device.
 
Event or Problem Description
The customer contacted physio-control to report that their device had all three icons (charge pak, attention and service wrench) present on its readiness display.The customer advised that the device did not provide any voice prompts when the device was powered on.The three icons being illuminated indicates that the device may not be able to provide sufficient defibrillation therapy, if necessary.There was no patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK CR(R) PLUS DEFIBRILLATOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
MDR Report Key7806197
Report Number0003015876-2018-01361
Device Sequence Number1142841
Product Code MKJ
UDI-Device Identifier00885074440323
UDI-Public00885074440323
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K033275
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCRPLUS
Device Catalogue Number99403-000156
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/17/2018
Initial Date Received by Manufacturer 08/01/2018
Supplement Date Received by Manufacturer09/04/2018
Initial Report FDA Received Date08/22/2018
Supplement Report FDA Received Date09/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-