ENCORE MEDICAL L.P. FOUNDATION HIP; HEAD, UNIPOLAR, MODULAR, ENCORE SIZE47 Back to Search Results
Model Number 411-00-147
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 08/20/2018
Type of Reportable Event Serious Injury
Event or Problem Description
Second revision surgery - due to the patient having an infection.The surgeon removed and replaced the unipolar and sleeve.
 
Additional Manufacturer Narrative
The reason for this revision surgery was reported as an infection.The previous surgery and the revision detailed in this investigation occurred 2.7 months apart.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the infection or inhibited the patient's immune system.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhrs) show that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to an infection.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The device was verified to have gone through an acceptable sterilization process and was within it's expiration date at the time of use during the previous surgery.There were no findings during this investigation that indicate that the reported device was the root cause or had a direct connection with the patient's infection.There are multiple factors that may contribute to an infection that are outside of the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOUNDATION HIP
Common Device Name
HEAD, UNIPOLAR, MODULAR, ENCORE SIZE47
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key7806348
Report Number1644408-2018-00810
Device Sequence Number1195630
Product Code KWL
UDI-Device Identifier00888912076821
UDI-Public(01)00888912076821
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K973614
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2022
Device Model Number411-00-147
Device Catalogue Number411-00-147
Device Lot Number914B1042
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/20/2018
Supplement Date Received by Manufacturer08/28/2018
Initial Report FDA Received Date08/22/2018
Supplement Report FDA Received Date08/29/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
411-00-350, LOT 931B1920
Outcome Attributed to Adverse Event Required Intervention;
Patient Age57 YR
-
-