Analysis: the samples were not returned from the user facility; therefore, device evaluations are unable to be performed.A lot history review revealed 4 reocclusion complaints with one associated with the same patient for this lot.A review of the device history record (dhr) indicates the lot was manufactured to specification.Conclusion: the actual samples were not received for evaluation.The dhr found nothing to indicate a manufacturing related cause for this event.The investigator assessed the event was possibly related to the study device, but not the procedure.Based on the instructions for use (ifu), the occurrence of reocclusion is an inherent risk of any pta procedure, and has been reported in clinical trials of drug coated balloons.If additional information becomes known to the manufacturer, a supplemental report will be provided with all relevant information.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
|
It was reported through a clinical registry that during the index procedure, three lutonix percutaneous transluminal drug coated balloon (dcb) dilatation catheters were used to treat the target lesion located in the left mid superficial femoral artery (sfa).Approximately 12 months after the index procedure, the patient¿s left mid sfa vasculature was reportedly reoccluded.No additional intervention has been performed at this time.Subsequently, 3 months later the hcp performed a target lesion reintervention and deemed the procedure successful.The investigator reassessed this event was possibly related to the study device, but not related to the procedure.The sample was discarded by the user facility and not available for further evaluation.No adverse patient effects were reported.This is one of three products involved with the reported event and the associated manufacturer¿s report numbers are 3006513822-2018-00180 and 3006513822-2018-00181.
|