AV-TEMECULA-CT XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Catalog Number 1550400-18
Medical Device Problem Code Unintended System Motion (1430)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/06/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The stent remains in the vessel; the delivery system was not returned to abbott vascular for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents reported from this lot.A cine was received and reviewed by an abbott clinical specialist.The reviewer concluded that the 4.0x18 sierra stent deployed in the proximal lad migrated proximally requiring additional percutaneous coronary intervention (pci) distally, where another stent was implanted.The investigation was unable to determine a conclusive cause for the reported stent movement.There is no indication of a product quality issue with respect to the design, manufacture or labeling of the device.
 
Event or Problem Description
It was reported that during a procedure of the proximal left anterior descending (lad) artery, during stent deployment, the 4.0 x 18 mm xience sierra stent moved proximally, resulting in the stent implanting and missing the distal portion of the lesion.A different 4.0 x 8 mm xience sierra stent was implanted in the distal lesion segment.Excellent angiographic result was achieved.There was no reported adverse patient effect or a clinically significant delay in procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7806580
Report Number2024168-2018-06498
Device Sequence Number11048607
Product Code NIQ
UDI-Device Identifier08717648227509
UDI-Public08717648227509
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date (Section B) 08/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/01/2019
Device Catalogue Number1550400-18
Device Lot Number8042041
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/02/2018
Initial Report FDA Received Date08/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age63 YR
Patient Weight98
-
-