|
Dexcom was made aware on (b)(4) 2018 that on (b)(6) 2018, the receiver had intermittent audio output.No additional event or patient information is available.The receiver was returned for evaluation.An exterior visual inspection was performed and passed.The receiver was charged and will boot.The receiver log was downloaded and reviewed, finding no errors related to the complaint.A receiver functional test was performed and passed all relevant testing.A global communication tool software test was performed and passed.Manual "try it" testing was performed and passed.The receiver case was opened for an interior inspection and passed.Speaker resistance was measured and was within specification.The reported event of an intermittent audio output was not confirmed.A probable cause could not be determined.
|