Catalog Number 328506 |
Device Problem
Difficult to Remove (1528)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/01/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Medical device expiration date: n/a.Investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.A review of the device history record was completed for batch# 7310516.All inspections were performed per the applicable operations qc specifications.There were two (2) notifications [(b)(4)] noted that did not pertain to the complaint.Investigation conclusion: based on the samples / photo(s) received the investigation concluded: unconfirmed: bd was not able to duplicate or confirm the customer¿s indicated failure as no samples or photos were returned.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.Root cause description: root cause cannot be determined at this time as the issue is unconfirmed as no samples or photos were returned.Rationale: based on the investigation, no additional investigation and no capa is required at this time.
|
|
Event Description
|
It was reported that the plunger on the bd ultra-fine¿ insulin syringe was difficult to move.There was no report of exposure, injury, or medical intervention.
|
|
Manufacturer Narrative
|
Correction: due to an it issue beginning on 7/3/2018, previously filed emdrs did not contain required fields.This supplemental emdr is filed to provide the following omitted fields: sex: male.Event attributed to: other.Device single use?: no.Device returned to manufacture: no.
|
|
Event Description
|
It was reported that the plunger on the bd ultra-fine¿ insulin syringe was difficult to move.There was no report of exposure, injury, or medical intervention.
|
|