MAUDE Adverse Event Report: BD MEDICAL - DIABETES CARE BD ULTRA-FINE¿ INSULIN SYRINGE; PISTON SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL - DIABETES CARE BD ULTRA-FINE¿ INSULIN SYRINGE; PISTON SYRINGE Back to Search Results
Catalog Number 328506
Device Problem Difficult to Remove (1528)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/01/2018
Event Type  malfunction  
Manufacturer Narrative
Medical device expiration date: n/a.Investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.A review of the device history record was completed for batch# 7310516.All inspections were performed per the applicable operations qc specifications.There were two (2) notifications [(b)(4)] noted that did not pertain to the complaint.Investigation conclusion: based on the samples / photo(s) received the investigation concluded: unconfirmed: bd was not able to duplicate or confirm the customer¿s indicated failure as no samples or photos were returned.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.Root cause description: root cause cannot be determined at this time as the issue is unconfirmed as no samples or photos were returned.Rationale: based on the investigation, no additional investigation and no capa is required at this time.
 
Event Description
It was reported that the plunger on the bd ultra-fine¿ insulin syringe was difficult to move.There was no report of exposure, injury, or medical intervention.
 
Manufacturer Narrative
Correction: due to an it issue beginning on 7/3/2018, previously filed emdrs did not contain required fields.This supplemental emdr is filed to provide the following omitted fields: sex: male.Event attributed to: other.Device single use?: no.Device returned to manufacture: no.
 
Event Description
It was reported that the plunger on the bd ultra-fine¿ insulin syringe was difficult to move.There was no report of exposure, injury, or medical intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ULTRA-FINE¿ INSULIN SYRINGE
Type of Device
PISTON SYRINGE
Manufacturer (Section D)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
MDR Report Key7806736
MDR Text Key118002476
Report Number1920898-2018-00642
Device Sequence Number1
Product Code FMF
UDI-Device Identifier00681131311748
UDI-Public00681131311748
Combination Product (y/n)N
PMA/PMN Number
K170386
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Type of Report Initial,Followup
Report Date 09/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/22/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number328506
Device Lot Number7310516
Was Device Available for Evaluation? No
Date Manufacturer Received08/03/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-