Model Number DTBB1D1
Medical Device Problem Codes High impedance (1291); Misconnection (1399)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/30/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that at the cardiac resynchronization therapy defibrillator (crt-d) change out procedure when the existing right ventricular (rv) lead was attached to the replacement crt-d the rv defibrillator impedance was high and undefined.The rv lead was removed and then reattached with the same results.Additionally, the rv and svc (superior vena cava) coils were swapped and the results in the rv coil header was still high and undefined impedance.A new crt-d was then given to the physician and it tested fine.The replacement crt-d was not used and the new crt-d was implanted instead.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the device was returned and analyzed.Returned product analysis was performed and no anomalies were found.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIVA S
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
MDR Report Key7806896
Report Number3004209178-2018-18914
Device Sequence Number1143522
Product Code NIK
UDI-Device Identifier00643169720329
UDI-Public00643169720329
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P010031
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/28/2019
Device Model NumberDTBB1D1
Device Catalogue NumberDTBB1D1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/06/2018
Initial Date Received by Manufacturer 07/30/2018
Supplement Date Received by Manufacturer09/05/2018
Initial Report FDA Received Date08/22/2018
Supplement Report FDA Received Date09/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age65 YR
Patient Weight107
-
-