| Model Number |
MMT-723RNAS |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/23/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Customer reported via phone call that the insulin pump alarmed compromised force sensor system, unexpected restart alarm and external ram crc error alarm on the insulin pump.The customer¿s blood glucose level was unknown.Troubleshooting was performed.Customer advised the insulin pump had multiple alarms and the issues remained unresolved.The customer was advised that the insulin pump needed to be replaced and was advised to discontinue use of the insulin pump and to revert to the backup plan.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed the displacement test.Device received compromised force sensor system alarm during the basic occlusion test due to loose or protruded drive support disk.Unable to perform the unexpected restart error test, prime test, excessive no delivery test and occlusion test or verify external ram crc error and unexpected restart error due to compromised force sensor system alarm.
|
| |
|