Model Number 9770500
Device Problem Poor Quality Image (1408)
Patient Problem No Code Available (3191)
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned to the manufacturer, at this time, for evaluation.A history review of serial number (b)(4) showed no other similar complaint(s) from this serial number.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device has not been returned to the manufacturer, at this time, for evaluation.
 
Event Description
Per clin spec and vat director, poor image quality.Unable to visualize the needle and complaining the image is gray and grainy.This is causing inaccuracy with placements which is costing the team additional money in product and causing the patients to have multiple sticks.
 
Event Description
Per clin spec and vat director, poor image quality.Unable to visualize the needle and complaining the image is gray and grainy.This is causing inaccuracy with placements which is costing the team additional money in product and causing the patients to have multiple sticks.
 
Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of poor image quality is unconfirmed.The scanner was evaluated and the image was found to be normal for a sr 8 scanner with the 32 mm probe.No other functionality issues with the equipment were found during evaluation/servicing.The device was serviced, tested, remanufactured and returned to the customer.A history review of serial number (b)(4) showed no other similar complaint(s) from this serial number.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE~RITE 8 ULTRASOUND SYSTEM (USA)
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7807789
MDR Text Key118229946
Report Number3006260740-2018-02195
Device Sequence Number1
Product Code IYO
UDI-Device Identifier00801741096891
UDI-Public(01)00801741096891
Combination Product (y/n)N
PMA/PMN Number
K152554
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 08/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9770500
Device Catalogue Number9770500
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2018
Event Location Hospital
Initial Date Manufacturer Received Not provided
Initial Date FDA Received08/22/2018
Supplement Dates Manufacturer Received08/13/2019
Supplement Dates FDA Received08/30/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-