Unit received with all operating currents within spec and passed functional testing including the rewind, basic occlusion test, occlusion test, prime/compromised force sensor system test, excessive no delivery test, self test, unexpected restart error test and displacement test.Pump passed the delivery accuracy test at 0.08720 inches.The drive support disk was inspected and no anomaly was noted.No excessive no delivery alarms, motor error alarms, external ram crc error alarms or unexpected pump resetting anomaly was noted.The motor was tested outside the device and passed.
|